EUR PET Luxuriant Veterinary Diagnostic Anesthesia Machine Comprehensive Vet Anesthesia Device Veterinary Instrument | ||
Product Parameters | ||
Ventilation mode | ||
IPPV, A/C, SIMV, SIGH, MANUAL | ||
Ventilator parameter range | ||
Flowmeter | O2 (0~5L/min) | |
Rapid oxygen supply | 25L/min~75L/min | |
Tidal volume (Vt) | 0, 20 mL ~ 1500 mL | |
Frequence(Freq) | 1/min~100/min | |
I:E | 2:1~1:6 | |
Pressure triggering sensitivity (Ptr) | -20 cmH2O~0 cmH2O(Based on PEEP) | |
Flow trigger sensitivity (Ftr) | 0.5 L/min ~ 30 L/min | |
SIGH | 0(off) 1/100 ~ 5/100 | |
Apnea ventilation | OFF,5 s ~ 60 s | |
Pressure Limit | 20 cmH2O ~ 100 cmH2O | |
Monitoring parameter | ||
Frequence(Freq) | 0 /min ~ 100 /min | |
Tidal volume (Vt) | 0 mL ~ 2000 mL | |
MV | 0 L/min ~ 100 L/min | |
Wooden case size | ||
Main component packing size | L 680*W 840*H 695mm | |
G.W. | 60KG | |
N.W. | 26KG | |
Accessary packing size | L 650*W 690*H 520mm | |
G.W. | 29KG | |
N.W. | 11KG | |
Total CBM | 0.633 | |







Alarm and protection | ||
The AC power failure alarm | Power failure or no connection | |
Low voltage alarm for back up battery | <11.3±0.3V | |
No tidal volume | ≤5mL within 6s | |
High oxygen concentration alarm | 19%-100% | |
Low oxygen concentration alarm | 18%-99% | |
High airway pressure alarm | 20cmH2O-100cmH2O | |
Low airway pressure alarm | 0cmH2O-20cmH2O | |
High minute volume alarm | 1L/min-20L/min | |
Low minute volume alarm | 0L/min-10L/min | |
Continuous pressure alarm | (PEEP+1.5kPa) over 16s | |
Suffocation warning | 5s-60s no spontaneous ventilation | |
The maximum limited pressure | <12.5 kPa | |
Fan error | Show on screen | |
Oxygen deficit | Show on screen | |
Working condition | ||
Gas source | O2 | |
Pressure | 280kPa-600kPa | |
Voltage | 100-240V | |
Power frequency | 50/60Hz | |
Oscillogram | ||
P-T (pressure – time) | ||
F-T (flow – time) | ||